Diagnosing and troubleshooting
10.2 General fault codes
SITRANS LR250 (PROFIBUS PA)
164 Operating Instructions, 01/2014, A5E32221386-AB
Code/
Icon
Meaning
Corrective Action
S: 9
Service interval as defined in
Reminder 2
(Demanded) (4.4.6.)
has expired.
Perform service.
S: 10
Input parameters
Low Calibration Point (1.6.)
and
High Calibration Point (1.7.)
are the same.
Check calibration settings of device.
Ensure settings for High Calibration Point and Low
Calibration Point are different.
S: 11
Internal temperature sensor failure. Repair required: contact your local Siemens
representative.
S: 12
Internal temperature of device has exceeded
specifications: it is operating outside its
temperature range.
Relocate device and/or lower process temperature
enough to cool device.
Inspect for heat-related damage and contact your
local Siemens representative if repair is required.
Fault code will persist until a manual reset is
performed using SIMATIC PDM or the LCD interface.
S: 14
Input Scaling (2.6.5.)
Upper and lower values
for AIFB1 are the same.
Check configuration for AIFB1.
Ensure that Upper Value and Lower Value (Input
Scaling) are not the same.
S: 15
Input Scaling (2.6.5.)
Upper and lower values
for AIFB2 are the same.
Check configuration for AIFB2.
Ensure that Upper Value and Lower Value (Input
Scaling) are not the same.
S: 17
Calibration interval as defined in
Reminder 1
(Required) (4.5.5.)
has expired.
Perform calibration.
S: 18
Calibration interval as defined in
Reminder 2
(Demanded) (4.5.6.)
has expired.
Perform calibration.
S: 28
* Internal device failure caused by a RAM
memory error.
Repair required: contact your local Siemens
representative.
S: 29
* EEPROM damaged. Repair required: contact your local Siemens
representative.
S: 31
* Flash error. Repair required: contact your local Siemens
representative.