User's Manual
Table Of Contents
- 8 Parameter reference
- 9 Service and maintenance
- 10 Diagnosing and troubleshooting
- 11 Technical data
- 12 Dimension drawings
- 12.1 Threaded horn antenna
- 12.2 Threaded horn antenna with extension
- 12.3 Flanged horn antenna
- 12.4 Flanged horn antenna with extension
- 12.5 Flanged encapsulated antenna (2"/DN50/50A sizes only)
- 12.6 Flanged encapsulated antenna (3"/DN80/80A sizes and larger)
- 12.7 Threaded PVDF antenna
- 12.8 Threaded connection markings
- 12.9 Raised-Face flange per EN 1092-1 for flanged horn antenna
- 12.10 Raised-Face flange per EN 1092-1 for flanged encapsulated antenna
- 12.11 Flat-Face flange
- 12.12 Process connection tag (pressure rated versions)
- A Appendix A: Technical reference
- B Appendix B: PROFIBUS PA profile structure
- C Appendix C: Communications via PROFIBUS
- D Appendix D: Certificates and Support
- 13 List of abbreviations
- 14 LCD menu structure
- Glossary
- Index
Diagnosing and troubleshooting
10.2 General fault codes
SITRANS LR250 (PROFIBUS PA)
164 Operating Instructions, 01/2014, A5E32221386-AB
Code/
Icon
Meaning
Corrective Action
S: 9
Service interval as defined in
Reminder 2
(Demanded) (4.4.6.)
has expired.
Perform service.
S: 10
Input parameters
Low Calibration Point (1.6.)
and
High Calibration Point (1.7.)
are the same.
• Check calibration settings of device.
• Ensure settings for High Calibration Point and Low
Calibration Point are different.
S: 11
Internal temperature sensor failure. Repair required: contact your local Siemens
representative.
S: 12
Internal temperature of device has exceeded
specifications: it is operating outside its
temperature range.
• Relocate device and/or lower process temperature
enough to cool device.
• Inspect for heat-related damage and contact your
local Siemens representative if repair is required.
• Fault code will persist until a manual reset is
performed using SIMATIC PDM or the LCD interface.
S: 14
Input Scaling (2.6.5.)
Upper and lower values
for AIFB1 are the same.
• Check configuration for AIFB1.
• Ensure that Upper Value and Lower Value (Input
Scaling) are not the same.
S: 15
Input Scaling (2.6.5.)
Upper and lower values
for AIFB2 are the same.
• Check configuration for AIFB2.
• Ensure that Upper Value and Lower Value (Input
Scaling) are not the same.
S: 17
Calibration interval as defined in
Reminder 1
(Required) (4.5.5.)
has expired.
Perform calibration.
S: 18
Calibration interval as defined in
Reminder 2
(Demanded) (4.5.6.)
has expired.
Perform calibration.
S: 28
* Internal device failure caused by a RAM
memory error.
Repair required: contact your local Siemens
representative.
S: 29
* EEPROM damaged. Repair required: contact your local Siemens
representative.
S: 31
* Flash error. Repair required: contact your local Siemens
representative.