User's Manual
Table Of Contents
- 8 Parameter reference
- 9 Service and maintenance
- 10 Diagnosing and troubleshooting
- 11 Technical data
- 12 Dimension drawings
- 12.1 Threaded horn antenna
- 12.2 Threaded horn antenna with extension
- 12.3 Flanged horn antenna
- 12.4 Flanged horn antenna with extension
- 12.5 Flanged encapsulated antenna (2"/DN50/50A sizes only)
- 12.6 Flanged encapsulated antenna (3"/DN80/80A sizes and larger)
- 12.7 Threaded PVDF antenna
- 12.8 Threaded connection markings
- 12.9 Raised-Face flange per EN 1092-1 for flanged horn antenna
- 12.10 Raised-Face flange per EN 1092-1 for flanged encapsulated antenna
- 12.11 Flat-Face flange
- 12.12 Process connection tag (pressure rated versions)
- A Appendix A: Technical reference
- B Appendix B: PROFIBUS PA profile structure
- C Appendix C: Communications via PROFIBUS
- D Appendix D: Certificates and Support
- 13 List of abbreviations
- 14 LCD menu structure
- Glossary
- Index
Parameter reference
SITRANS LR250 (PROFIBUS PA)
Operating Instructions, 01/2014, A5E32221386-AB
131
Event Diagnosis (3.5.3.)
Allows you to assign one of the diagnostic options listed below to any of the events listed in
Event Index (3.5.1.)
. This allows you to tailor a device response appropriate for your particular
process.(Event Diagnosis affects Condensed Acyclic Diagnostics and Cyclic Extended
Diagnostics). See
Condensed Mode Diagnosis (Page 239) for more detail.
Event Diagnosis Options
Status OK
Maintenance Required
Maintenance Demanded
*
Maintenance alarm
Invalid process conditions
Function check or simulation
To assign a diagnosis to a particular event via the handheld programmer:
1. Go to
Enable (3.4.1.)
and select
Yes
to enable Condensed Mode.
2. Go to
Event Index (3.5.1.)
and enter the event index number corresponding to a particular
event.
3. Go to
Event Diagnosis (3.5.3.)
and choose a Diagnosis option from the table above.
To assign a status to a particular event via SIMATIC PDM:
1. Go to
Level Meter > Diagnostics > Condensed Status Setup
, and select
Yes
to enable
Condensed Status Mode.
2. Go to
Level Meter > Diagnostics > Condensed Status
.
3. Select the Diagnosis line for the selected Event, then choose a Diagnosis option from the
associated pull-down menu.
Peak Values (3.6.)
To view via SIMATIC PDM:
Open the menu
View – Device Diagnostics
, select
Device Status
, and click on the tab
Device
Status
. For more details see Device Diagnostics (Page 86).
Min. Measured Value (3.6.1.)
The minimum recorded Sensor value, reported in units defined in
Unit (2.3.1.)
.
Max. Measured Value (3.6.2.)
The maximum recorded Sensor value, reported in units defined in
Unit (2.3.1.)
.