User's Manual
Table Of Contents
- 8 Parameter reference
- 9 Service and maintenance
- 10 Diagnosing and troubleshooting
- 11 Technical data
- 12 Dimension drawings
- 12.1 Threaded horn antenna
- 12.2 Threaded horn antenna with extension
- 12.3 Flanged horn antenna
- 12.4 Flanged horn antenna with extension
- 12.5 Flanged encapsulated antenna (2"/DN50/50A sizes only)
- 12.6 Flanged encapsulated antenna (3"/DN80/80A sizes and larger)
- 12.7 Threaded PVDF antenna
- 12.8 Threaded connection markings
- 12.9 Raised-Face flange per EN 1092-1 for flanged horn antenna
- 12.10 Raised-Face flange per EN 1092-1 for flanged encapsulated antenna
- 12.11 Flat-Face flange
- 12.12 Process connection tag (pressure rated versions)
- A Appendix A: Technical reference
- B Appendix B: PROFIBUS PA profile structure
- C Appendix C: Communications via PROFIBUS
- D Appendix D: Certificates and Support
- 13 List of abbreviations
- 14 LCD menu structure
- Glossary
- Index
Parameter reference
SITRANS LR250 (PROFIBUS PA)
Operating Instructions, 01/2014, A5E32221386-AB
129
Event Index (3.5.1.)
The numeric component of the Event Code for a Condensed Status event. Use the index
number to identify a particular event in the list below.
Event index
Event code
Event description
a)
0
S0
Loss of Echo
2
S2
No Tech Power
10
S10
Level Transducer Block (LTB) Scale
11
S11
Internal Temperature Sensor
12
S12
Internal Temperature High
14
S14
AIFB1 PV Range
15
S15
AIFB2 PV Range
28
S28
Memory RAM
29
S29
Memory EEPROM
30
S30
Memory EEPROM Flags
31
S31
Memory Flash
33
S33
Internal Temperature Calibration
34
S34
Velocity Calibration
35
S35
Receiver Init Calibration
36
S36
Receiver Calibration
37
S37
Technology Module Calibration
38
S38
Technology Module Ramp
a)
See General Fault Codes (Page 163) for the meaning of each event.
For example:
Event Code for Loss of Echo
=
S0
Event Index
=
0
To select a particular event via the handheld programmer:
1. Go to
Enable (3.4.1.)
and select
Yes
to enable Condensed Mode.
2. Go to
Event Index (3.5.1.)
and enter the event index number corresponding to the event.
To select a particular event via SIMATIC PDM:
1. Go to
Diagnostics > Condensed Status Setup > Condensed Status Mode
and select
Yes
to
enable
Condensed Mode.
2. Go to
Diagnostics > Condensed Status
.
3. For each event, you can select either the Status or the Diagnosis line, then choose a Status
or Diagnosis option from the associated pull-down menu.