Appendix C: Communications via PROFIBUS
C.10 Diagnostics
SITRANS LR250 (PROFIBUS PA)
Operating Instructions, 01/2014, A5E32221386-AB
243
Acyclic Extended Diagnostics /General Fault Codes
LCD display
Meaning
Corrective Action
Byte
Bit
S:32
IDENT number conflict. Ensure value of the Ident number selector
is correct for the network configuration. If it
is correct, the device needs to be re-
parameterized by the PLC.
4
0
S:33
Factory calibration for the internal
temperature sensor has been lost.
Repair required: contact your local
Siemens representative.
1
S:34
Factory calibration for the device has
been lost.
Repair required: contact your local
Siemens representative.
2
S:35
Factory calibration for the device has
been lost.
Repair required: contact your local
Siemens representative.
3
S:36
Unable to start microwave module. Cycle power. If fault persists, contact your
local Siemens representative.
4
S:37
Measurement hardware problem. Cycle power. If fault persists, contact your
local Siemens representative.
5
S:38
Microwave module hardware failure:
unable to calculate distance
measurement.
Cycle power. If fault persists, contact your
local Siemens representative.
6
S:43
Factory calibration for the radar receiver
has been lost.
Repair required: contact your local
Siemens representative.
5 3