User's Manual
Table Of Contents
- 8 Parameter reference
- 9 Service and maintenance
- 10 Diagnosing and troubleshooting
- 11 Technical data
- 12 Dimension drawings
- 12.1 Threaded horn antenna
- 12.2 Threaded horn antenna with extension
- 12.3 Flanged horn antenna
- 12.4 Flanged horn antenna with extension
- 12.5 Flanged encapsulated antenna (2"/DN50/50A sizes only)
- 12.6 Flanged encapsulated antenna (3"/DN80/80A sizes and larger)
- 12.7 Threaded PVDF antenna
- 12.8 Threaded connection markings
- 12.9 Raised-Face flange per EN 1092-1 for flanged horn antenna
- 12.10 Raised-Face flange per EN 1092-1 for flanged encapsulated antenna
- 12.11 Flat-Face flange
- 12.12 Process connection tag (pressure rated versions)
- A Appendix A: Technical reference
- B Appendix B: PROFIBUS PA profile structure
- C Appendix C: Communications via PROFIBUS
- D Appendix D: Certificates and Support
- 13 List of abbreviations
- 14 LCD menu structure
- Glossary
- Index
Parameter reference
SITRANS LR250 (PROFIBUS PA)
Operating Instructions, 01/2014, A5E32221386-AB
107
Entering breakpoints via the hand-held programmer:
1. The default for level values is percent: if you want to select units instead, navigate to
Setup
(2.) > Sensor (2.3.) > Level Unit (2.3.2.)
, and select the desired unit.
2. Navigate to
Setup (2.) > Sensor (2.3.) > PV Units (volume/level) (2.3.3.)
, and select the
desired volume units.
3. Go to
XY index (2.4.1.5.)
and enter the number of the breakpoint you wish to adjust: for
example, for breakpoint
1
enter
1
.
4. Go to
X value (2.4.1.6.)
and enter the level value for the breakpoint just identified.
5. Go to
Y value (2.4.1.7.)
and enter the volume value for the breakpoint just identified.
6. Repeat steps 3 to 5 until values have been entered for all required breakpoints.
X value (2.4.1.6.)
Y value (2.4.1.7.)