Appendix C: Communications via PROFIBUS
C.10 Diagnostics
SITRANS LR250 (PROFIBUS PA)
Operating Instructions, 01/2014, A5E32221386-AB
241
Acyclic Extended Diagnostics /General Fault Codes
LCD display
Meaning
Corrective Action
Byte
Bit
S:0
The device was unable to get a
measurement within the Failsafe LOE
Timer period. Possible causes: faulty
installation, antenna material buildup,
foaming/other adverse process
conditions, invalid calibration range.
Ensure installation details are correct.
Ensure no antenna material
buildup. Clean if necessary.
Adjust process conditions to minimize
foam or other adverse conditions.
Correct range calibration.
If fault persists, contact your local
Siemens representative.
0
0
S:2
Unable to collect profile because of a
power condition that is outside the
operating range of the device.
Repair required. Contact your local
Siemens representative.
2
S:3
Device is nearing its lifetime limit
according to the value set in
Maintenance Required Limit.
Replacement is recommended. 3
S:4
Device is nearing its lifetime limit
according to the value set in
Maintenance Demanded Limit.
Replacement is recommended. 4
S:6
Sensor is nearing its lifetime limit
according to the value set in
Maintenance Required Limit.
Replacement is recommended. 6
S:7
Sensor is nearing its lifetime limit
according to the value set in
Maintenance Demanded Limit.
Replacement is recommended. 7
S:8
Service interval as defined in
Maintenance Required Limit has expired.
Perform service
0
S:9
Service interval as defined in
Maintenance Demanded Limit
has expired.
Perform service. 1
S:10
Input parameters High Calibration Point
and Low Calibration Point are the same.
Check calibration settings of device.
Ensure settings for High Calibration
Point and Low Calibration Point are
different.
3
S:11
Internal temperature sensor failure. Repair required. Contact your local
Siemens representative.
4