Appendix A: Technical reference
A.2 Echo processing
SITRANS LR250 (PROFIBUS PA)
Operating Instructions, 01/2014, A5E32221386-AB
205
Example: CLEF off: Position set to Hybrid
Vessel height: 1.5 m; CLEF range set to 0 (
Center
algorithm gives the same result.)
①
default TVT
②
material echo
③
vessel bottom echo selected
④
echo marker
Example: CLEF enabled
Vessel height: 1.5 m; CLEF range set to 0.5 m
①
default TVT
②
material echo selected
③
vessel bottom echo
④
echo marker