10.2 General fault codes
SITRANS LR250 (FOUNDATION FIELDBUS)
Operating Instructions, 01/2014, A5E32221411-AB
191
Code/
Icon
Meaning
Corrective Action
S: 10
Input parameters
Low Calibration Point (2.3.7.1.)
and
High Calibration Point (2.3.7.2.)
are the
same.
Check calibration settings of device.
Ensure settings for High Calibration Point and Low
Calibration Point are different.
S: 11
Internal temperature sensor failure. Repair required: contact your local Siemens representative.
S: 12
Internal temperature of device has exceeded
specifications: it is operating outside its
temperature range.
Relocate device and/or lower process temperature
enough to cool device.
Inspect for heat-related damage and contact your local
Siemens representative if repair is required.
Fault code will persist until a manual reset is performed
using AMS or the LCD interface.
S: 17
Calibration interval (4.5.1.)
has
expired and has triggered a
Maintenance
Required reminder (4.5.5.)
.
Perform calibration.
S: 18
Calibration interval (4.5.1.)
has
expired and has triggered a
Maintenance
Demanded reminder (4.5.6.)
.
Perform calibration.
S: 22
Time Base Fault Hardware problem: contact your local Siemens
representative.
S: 28
Internal device failure caused by a RAM memory
error.
Repair required: contact your local Siemens representative.
S: 29
EEPROM damaged. Repair required: contact your local Siemens representative
S: 31
Flash error. Repair required: contact your local Siemens representative
S: 33
Factory calibration for the internal temperature
sensor has been lost.
Repair required: contact your local Siemens representative
S: 34
Factory calibration for the device has been lost. Repair required: contact your local Siemens representative