Parameter reference
SITRANS LR250 (FOUNDATION FIELDBUS)
130 Operating Instructions, 01/2014, A5E32221411-AB
Quick Start (1.)
Note
For detailed instructions see
Quick Start Wizard via the handheld programmer (Page 48) or
Quick Start Wizard via AMS Device Manager
(Page 70).
Setup (2.)
Note
See Parameter menus (Page 42) or Operating via AMS Device Manager (Page 63) for
instructions.
Default settings in the parameter tables are indicated with an asterisk (*) unless explicitly
stated.
Values shown in the following tables can be entered via the handheld programmer.
Identification (2.1.)
Tag (2.1.1.)
Read only. Text that can be used in any way. A recommended use is as a unique label for a field
device in a plant. Limited to 32 ASCII characters.
Note
The tag can only be changed from a remote master such as NIFBUS
-Configurator or DeltaV.
Descriptor (2.1.2.)
Read only. Text that can be used in any way. Limited to 32 ASCII characters. No specific
recommended use.
To access this parameter via AMS Device Manager see
Identification
under Identification
(RESOURCE) (Page 97).