User's Manual

Table Of Contents
Remote operation
7.1 Operating via SIMATIC PDM
SITRANS LR250 (PROFIBUS PA)
78 Operating Instructions, 01/2014, A5E32221386-AB
To set Device/Sensor Maintenance schedules:
1. Open the menu
Device Maintenance
, and click on the
Remaining Device/Sensor Lifetime
tab.
2. Modify desired values, and if desired, set reminders for either or both of
Reminder 1 before
Lifetime (Required)
/
Reminder 2 before Lifetime (Demanded)
.
3. Click
Write.
4. Click
Read
, to see the effects of your modification.
5. Click
Snooze
to add a year to the Total Expected Device Life.
To set Service/Calibration schedules:
1. Open the menu
Device Maintenance
, and click on the
Service/Calibration Schedule
tab.
2. Modify desired values and if desired, set reminders for either or both of
Reminder 1 before
Lifetime (Required)
/
Reminder 2 before Lifetime (Demanded)
.
3. Click
Write.
4. Click
Read
, to see the effects of your modification.
5. Click
Service/Calibration Performed
to reset the schedule.
7.1.5.5
Acknowledge Faults
Open the menu
Device Acknowledge Faults
, select the appropriate item from the Extended
Diagnostics pull-down menu, and click on
Transfer
.
7.1.5.6
Wear
Reports the number of hours the device has been operating, and the number of times it has
been powered up.
Open the menu
Device Wear
to view:
Powered Hours
Power-on Resets