User's Manual

Table Of Contents
Commissioning
6.2 Application examples
SITRANS LR250 (FOUNDATION FIELDBUS)
58 Operating Instructions, 01/2014, A5E32221411-AB
6.2.2
Horizontal vessel with volume measurement
Note
The minimum distance from the flange face to the target is lim
ited by
Near Range (2.5.1.)
.
To obtain level measurement proportional to vessel volume in a chemical vessel:
Low Calibration Point = 3.5 m (11.48 ft) from sensor reference point
High Calibration Point = 0.5 m (1.64 ft) from sensor reference point
Max. fill/empty rate = 0.2 m/min (0.65 ft/min)
Select vessel shape, Parabolic Ends, and enter values for A and L, to obtain a volume
reading instead of level.
In the event of a loss of echo:
SITRANS LR250 is to report a status of BAD or UNCERTAIN after 120 seconds (2 minutes).
Sensor reference point
A = 0.8 m
0.5 m
L = 6 m
High calibration point
100% = 8000 L
3.5. m
Volume reading
Low calibration point
Parameter type
Parameter Name
Options/
Values
Function
Quick Start
Wizard
parameters
Introduction
NEXT
Continue with Wizard.
Language
NEXT
Continue with current language.
Material
LIQUID
Response Rate
MED
Medium =1 m/minute
Units
M
meters
Low Calibration Point
3.5
3.5 m (11.48 ft)
High Calibration Point
0.5
0.5 m (1.64 ft)
Wizard Complete
FINISH
Transfers Quick Start settings to
device.