7.1 Operating via AMS Device Manager
SITRANS LR250 (FOUNDATION FIELDBUS)
114 Operating Instructions, 01/2014, A5E32221411-AB
7.1.17
Process Variables (Level Transducer Block - LTB)
To compare outputs in real time navigate to
Process Variables > LTB
.
Click on
Process Variables
then the
Values
tab to read the following:
Primary Variable
[see
Main Output (PV - Primary Value) (2.8.1.)
]
Level/Volume (PV)
The primary variable and the channel 1 output from the transducer block. For level
applications, chart range is affected by High and Low Level Point values set in
Configure/Setup > LTB > Setup > Sensor
. For volume applications, chart range is 0 to Max.
Volume, set in
Configure/Setup > LTB > Setup > Linearization
.
Click on
Trend View
tab to read the following:
Trend Values
Level/Volume (PV)
The primary variable and the channel 1 output from the transducer block.
Click on
Echo Profile
to open the dialog window to read the following: