Page 82 IQ Radar 300 PL-611
Communications
Communications
Communications
Communications
The measurement registers is 41,010.
The available register
Data Registers Parameter
Reading 41,010 P920
Volume 41,020 P924
Temperature 41,030 and 41,031 P912
The reading is expressed as a percentage of full scale, multiplied by 100:
Reading Value
0 0.00%
5000 50.00%
7564 75.64%
20,000 200.00%
Input/Output (R41,070 R41,143)
The IQ 300 has one mA output.
mA Output (R41,110)
The mA output is scaled from 0 to 2,000 (0 to 20mA multiplied by 100). Displayed in P911.
Parameter Access (R43,997 R46,999)
Parameter values are given as integers in the range of registers from R44,000 to R44,999.
The last three numbers of the register correspond to the parameter number.
Parameter Register # Format Register # Parameter #
44,000 46,000 P000
44,001 46,001 P001
44,002 46002 P002
……
44,999 46,999 P999
The parameters are usually all read/write.
Note:
Parameters P000 and P999 are read only. If P000 is set to LOCK ON then all of
the parameters are read only.
Parameter P999 (Master Reset) cannot be used via Modbus.
See Data Types on page 85 for a description of the different types of data
associated with different parameters.
However, before a parameter can be read or written to, the format (where decimal place
is) and the indexes must be defined.