PL-611 IQ Radar 300 Page 79
Parameters
Parameters
Parameters
Parameters
P904 Keypad
Press each keypad key in the following sequence:
As each key is pressed, the associated keypad number is displayed. On successful test
completion, the display reads PASS
PASSPASS
PASS. The display reads FAIL if a key is pressed out of
sequence or the programmer keypad malfunctions.
P911 mA Output Value
Displays the current value of the mA output in MilliAmps.
Values Range: 0.00 to 20.00
When P201 is set to 0 (manual), a test value can be entered and the displayed value is
transmitted to the output. Upon returning to the run mode, this value will remain. For all
other mA output functions, this parameter is read only.
P920 Reading Measurement
Displays the reading measurement that the unit is programmed for in RUN mode (P001,
operation).
Values (view only)
units showing Level/Space/Distance
P921 Material Measurement
Displays the reading measurement as though the unit were programmed to read Level
(P001 = 1).
P922 Space Measurement
Displays the reading measurement as though the unit were programmed to read Space
(P001 = 2).
P923 Distance Measurement
Displays the reading measurement as though the unit were programmed to read
Distance (P001 = 3).
P924 Volume Measurement
The calculated vessel capacity in Max Volume (P051) or % of Max Volume.