User's Manual
Table Of Contents
- Table of Contents
- General Information
- Specifications
- Installation
- Location
- Dimensions: IQ Radar 300 with Rod Antenna
- Dimensions: Threaded Rod
- Dimensions: Horn
- Mounting
- Rod Assembly
- Mounting: Rod Assembly
- Mounting: Manhole Covers
- Mounting: Horn Antennas
- Mounting: Waveguide Antenna
- Mounting: Stillpipe or Sidepipe
- Mounting: Horn with Waveguide Extensions
- Mounting: Sanitary Mounting
- Mounting: Location
- Interconnection
- Start Up
- Operation(
- Application Examples
- Parameter Descriptions
- _
- _
- Quick Start Parameters (P001 to P007)
- Volume Parameters (P050 to P053)
- Display and Reading Parameters (P060 to P063)
- Fail-Safe Parameters (P070 to P072)
- mA Parameters (P200 to P219)
- Independent mA Setpoints Parameters (P210 and P211)
- Profile Record Parameters (P330 to P337)
- Auto Record ON and OFF Setpoint Parameters (P334 to P337)
- Installation Records Parameter
- Range Calibration Parameters
- Rate Parameters
- Measurement Verification Parameters
- Communication Parameters
- Echo Processing Parameters
- Algorithm Parameters
- TVT Adjustment Parameters
- P837 Auto Near TVT
- Test Parameters
- _
- Communications: Modbus Register Map
- Communications: Data Types
- Error Handling
- Troubleshooting
- Maintenance(
- Appendix I
- Appendix II
- Appendix III
- Appendix IV
- Appendix V
PL-611 IQ Radar 300 Page 79
Parameters
Parameters
Parameters
Parameters
P904 Keypad
Press each keypad key in the following sequence:
As each key is pressed, the associated keypad number is displayed. On successful test
completion, the display reads PASS
PASSPASS
PASS. The display reads FAIL if a key is pressed out of
sequence or the programmer keypad malfunctions.
P911 mA Output Value
Displays the current value of the mA output in MilliAmps.
Values Range: 0.00 to 20.00
When P201 is set to 0 (manual), a test value can be entered and the displayed value is
transmitted to the output. Upon returning to the run mode, this value will remain. For all
other mA output functions, this parameter is read only.
P920 Reading Measurement
Displays the reading measurement that the unit is programmed for in RUN mode (P001,
operation).
Values (view only)
units showing Level/Space/Distance
P921 Material Measurement
Displays the reading measurement as though the unit were programmed to read Level
(P001 = 1).
P922 Space Measurement
Displays the reading measurement as though the unit were programmed to read Space
(P001 = 2).
P923 Distance Measurement
Displays the reading measurement as though the unit were programmed to read
Distance (P001 = 3).
P924 Volume Measurement
The calculated vessel capacity in Max Volume (P051) or % of Max Volume.