User's Manual

Table Of Contents
Page 76 IQ Radar 300 PL-611
Parameters
Parameters
Parameters
Parameters
TVT Adjustment Parameters
The following parameters are for authorized Milltronics Service personnel or
Technicians familiar with Milltronics echo processing techniques.
P830 TVT Type
Selects the TVT Curve used.
1
reserved
2
reserved
3
reserved
4
Smooth 2 (Factory Set)
5
reserved
Values
6
reserved
Related Parameters
P835 TVT Slope Min
Select the TVT type which gives the highest confidence (P805) under all level conditions.
Use this parameter with caution, and do not use TVT Slopes with the bF or bLF
Algorithm (P820).
P831 TVT Shaper
Turns the TVT Shaper ON
ONON
ON or OFF
OFFOFF
OFF.
0
00
0off
Values
1
11
1on
Related Parameters
P832 TVT Shaper Adjust
Turn the TVT Shaper ON before using P832, and afterwards turn the TVT Shaper ON and
OFF while monitoring the effect to pick up the true echo.
P832 TVT Shaper Adjust
Allows manual adjustment of the TVT curve.
Range: -50 to 50
Values
Pre-set: 0
Related Parameters
P831 TVT Shaper
Use this feature to bias the shape of the TVT curve to avoid crossing false echoes from
fixed objects.
Adjustment to this parameter is best done while viewing the echo profile with Dolphin
Plus. Refer to the Dolphin Plus online help for details. The TVT curve is divided into 40
breakpoints, accessible by enabling the point number as the breakpoint index field. Each
breakpoint is normalized to a value of 0, as displayed in the parameter value field. By
changing the breakpoint value, up or down, the intensity of the bias applied to that
breakpoint of the curve is respectively changed. By changing the value of adjacent