P705 Rate Update Time
Sets the time period (in seconds) over which the material level rate of change is
averaged before Rate Value update.
Values Range: 0.000 to 9999
Related Parameter
P707 Rate Value
P706 Rate Update Distance
The material level change (in metres) to initiate a Rate Value update.
Values Range: 0.000 to 9999
Related Parameter
P707 Rate Value
P707 Rate Value
The rate of material level change (in Units (P005) or % of Span (P007) per minute).
Values (view only) Display: -999 to 9999
Altered By
P704 Rate Filler
Related Parameters
P005 Units
P007 Span
A negative rate indicates the vessel is emptying.
Measurement Verification Parameters
P710 Fuzz Filter
Stabilizes the reported level due to level fluctuations (such as a rippling or splashing
liquid surface) within the Echo Lock Window (P713).
Values Range: 0 to 100 (0= off)
Altered by
P003 Measurement Response
Related Parameters
P007 Span
P713 Echo Lock Window
This value (in % of Span, P007) is automatically altered when Measurement Response
Speed (P003) changes. The higher the value entered, the greater the range of stabilized
fluctuation.