PL-611 IQ Radar 300 Page 69
Parameters
Parameters
Parameters
Parameters
P003 Value Meters/Minute
10.1
21
310
4 100
5 1000
P701 Maximum Empty Rate
Adjusts the IQ 300 response to decreases in the actual material level (or an advance to a
lower Failsafe Material Level, P071).
Values
ValuesValues
Values
Range: 0.000 to 9999 (stored in meters)
Altered by
Altered byAltered by
Altered by
P003 Measurement Response
Related Parameters
Related ParametersRelated Parameters
Related Parameters
P005 Units
P007 Span
P071 Fail-Safe Material Level
Enter a value slightly greater than the vessels maximum emptying rate. This value, in
Units (P005) or % of Span (P007) per minute, is automatically altered when Measurement
Response Speed (P003) is altered. Any emptying rate above this value will trigger any
alarms set to Rate.
P003 Value Meters/Minute
10.1
21
310
4 100
5 1000
P704 Rate Filter
Damps Rate Value (P707) fluctuations.
0
rate display not required
1
continuously filtered and updated
Values
2
update rate (use P705/P706)
Alters
P707 Rate Value
Related Parameters
P705 Rate Update Time
P706 Rate Update Distance
Enter the time or distance interval over which the Rate Value is to be calculated before
the display updates.
This is automatically altered along with Measurement Response Speed (P003).