Parameters
Parameters
Parameters
Parameters
Pipe Size (i.d.) Propagation Factor
50mm (2) 0.827
80mm (3) 0.915
100mm (4) 0.955
150mm (6) 0.980
20 mm (8) 0.990
Consult the factory for other sizes and propagation factor numbers.[Tim: who at the
factory?]
Note:
: :
: For waveguide antennas used as stillpipes, this value is on the product tag.
The propagation factor is constant for a given
pipe diameter,
or can be determined by
comparing the radar distance reading to the
actual process material distance (measured
from the face of the IQ 300 flange).
actual distance
= p.f.
IQ 300 distance
Using the readings shown:
10.42m
= 0.827
12.6m
Enter the propagation factor: 0.827
0.8270.827
0.827
Rate Parameters
These parameters determine how material level changes are reported.
P700 Maximum Fill Rate
Adjusts the IQ 300 response to increases in the actual material level (or an advance to a
higher Failsafe Material Level, P071).
Values
ValuesValues
Values
Range: 0.000 to 9999 (stored in meters)
Altered by
Altered byAltered by
Altered by
P003 Measurement Response
Related Parameters
Related ParametersRelated Parameters
Related Parameters
P005 Units
P007 Span
P071 Fail-Safe Material Level
Enter a value slightly greater than the maximum vessel filling rate. This value, in Units
(P005) or % of Span (P007) per minute, is automatically altered when Maximum Process
Speed (P003) is altered.