- Quick Start Parameters (P001 to P007)
- Volume Parameters (P050 to P053)
- Display and Reading Parameters (P060 to P063)
- Fail-Safe Parameters (P070 to P072)
- mA Parameters (P200 to P219)
- Independent mA Setpoints Parameters (P210 and P211)
- Profile Record Parameters (P330 to P337)
- Auto Record ON and OFF Setpoint Parameters (P334 to P337)
- Installation Records Parameter
- Range Calibration Parameters
- Rate Parameters
- Measurement Verification Parameters
- Communication Parameters
- Echo Processing Parameters
- Algorithm Parameters
- TVT Adjustment Parameters
- P837 Auto Near TVT
- Test Parameters
- _
- Communications: Modbus Register Map
- Communications: Data Types
- Error Handling
- Troubleshooting
- Maintenance(
- Appendix I
- Appendix II
- Appendix III
- Appendix IV
- Appendix V
PL-611 IQ Radar 300 Page 57
Parameters
Parameters
Parameters
Parameters
P005 Units
Specifies units for programming and measurement.
1
metres
2
centimetres
3
millimetres
4
feet
Values
5
inches
P006 Empty
Sets the distance in units (see P005) from flange face
to empty level.
Values -----
# units set in P005
Empty level can be set at any distance
desired, not just actual empty.
P007 Span
Sets the distance from empty (P006) to full/100% level.
Values -----
# units set in P005
Full level can be set at any measurement
above the empty level.
Volume Parameters (P050 to P053)
Sets the IQ Radar 300 to calculate readings based on reservoir volume, rather than level
P050 Tank Shape
Enter the Tank Shape value that matches the monitored vessel or reservoir.
When Operation is LEVEL
LEVELLEVEL
LEVEL (P001 = 1), liquid (material) volume is calculated. When
operation is SPACE
SPACESPACE
SPACE, remaining vessel capacity is calculated.
In the RUN
RUNRUN
RUN
mode, Readings are displayed in volumetric units, see Maximum Volume
(P051) on page 59. When percent is selected, the displayed volume Reading is a
percentage of Maximum Volume.
Note: Parameters P052 and P053 set tank dimensions A and L.