PL-611 IQ Radar 300 Page 53
Applications
Applications
Applications
Applications
Application Example: Stillpipe
An alternate to the waveguide antenna, this option is also used for products with an ε
r
or
less than 3 or if extremely turbulent or vortex conditions exist.
Note: For ε
r
< 3, the lower 40cm of vessel level may not be measurable.
This mounting arrangement can also be used to provide optimum signal conditions on
foaming materials.
Suitable pipe diameters are 50mm (2) to 250mm (10).
Note: The measurement range is reduced to 13m (43) for a 50mm (2) stillpipe or
14.5m (47) for an 80mm (3) stillpipe.
Parameter Enter
P999
----
master reset
P001
1
mode of measurement = level
P002
1
material = liquid
P003
2
measurement response = 1m/minute
P004
240
antenna
= dielectric rod, standard
length
P005
1
units = metres
P006
5
empty distance = 5m
P007
4.5
span = 4.5m
P655 (see
chart below)
0.955
propagation factor = 10mm (4) pipe
P820
6
algorithm = factory
P830
4
TVT type = factory
P837
(Note below)
2 & 1
Auto Near TVT = 0 (factory set)
vent hole
100mm (4) diameter