User's Manual
Table Of Contents
- Table of Contents
- General Information
- Specifications
- Installation
- Location
- Dimensions: IQ Radar 300 with Rod Antenna
- Dimensions: Threaded Rod
- Dimensions: Horn
- Mounting
- Rod Assembly
- Mounting: Rod Assembly
- Mounting: Manhole Covers
- Mounting: Horn Antennas
- Mounting: Waveguide Antenna
- Mounting: Stillpipe or Sidepipe
- Mounting: Horn with Waveguide Extensions
- Mounting: Sanitary Mounting
- Mounting: Location
- Interconnection
- Start Up
- Operation(
- Application Examples
- Parameter Descriptions
- _
- _
- Quick Start Parameters (P001 to P007)
- Volume Parameters (P050 to P053)
- Display and Reading Parameters (P060 to P063)
- Fail-Safe Parameters (P070 to P072)
- mA Parameters (P200 to P219)
- Independent mA Setpoints Parameters (P210 and P211)
- Profile Record Parameters (P330 to P337)
- Auto Record ON and OFF Setpoint Parameters (P334 to P337)
- Installation Records Parameter
- Range Calibration Parameters
- Rate Parameters
- Measurement Verification Parameters
- Communication Parameters
- Echo Processing Parameters
- Algorithm Parameters
- TVT Adjustment Parameters
- P837 Auto Near TVT
- Test Parameters
- _
- Communications: Modbus Register Map
- Communications: Data Types
- Error Handling
- Troubleshooting
- Maintenance(
- Appendix I
- Appendix II
- Appendix III
- Appendix IV
- Appendix V
Page 52 IQ Radar 300 PL-611
Applications
Applications
Applications
Applications
Application Example: Sliding Waveguide on
Anaerobic Digesters
Blanking, P800, and offset P652 are set at the factory. Check the device tag for specific
values.
Maximum level should be maintained at
least .46m (18”) from the end of the horn.
The raised position is for installation and
maintenance. The lowered position is for
operation. Program the unit for operation in
the lowered position.
Parameter Enter
P999
----
master reset
P001
1
mode of measurement = level
P002
1
material = liquid
P003
2
measurement response = 1m/minute
P004
240
antenna
= dielectric rod, standard
length
P005
1
units = metres
P006
5
empty distance = 5m
P007
4.5
span = 4.5m
P652
tag value
offset correction = factory
P800
tag value
blanking = factory
P820
1
algorithm = factory
P830
4
TVT type = factory
P837
(Note below)
2 & 1
Auto Near TVT = 0 (factory set)
Note: Only set P837 if product is at least 2.5m (78”) away. If closer, leave P837 at 0
00
0
until such time the level drops below 2.5m (78”).
Run: To start normal operation, press the PROGRAM key .
P006
P007
horn