User's Manual
Table Of Contents
- Table of Contents
- General Information
- Specifications
- Installation
- Location
- Dimensions: IQ Radar 300 with Rod Antenna
- Dimensions: Threaded Rod
- Dimensions: Horn
- Mounting
- Rod Assembly
- Mounting: Rod Assembly
- Mounting: Manhole Covers
- Mounting: Horn Antennas
- Mounting: Waveguide Antenna
- Mounting: Stillpipe or Sidepipe
- Mounting: Horn with Waveguide Extensions
- Mounting: Sanitary Mounting
- Mounting: Location
- Interconnection
- Start Up
- Operation(
- Application Examples
- Parameter Descriptions
- _
- _
- Quick Start Parameters (P001 to P007)
- Volume Parameters (P050 to P053)
- Display and Reading Parameters (P060 to P063)
- Fail-Safe Parameters (P070 to P072)
- mA Parameters (P200 to P219)
- Independent mA Setpoints Parameters (P210 and P211)
- Profile Record Parameters (P330 to P337)
- Auto Record ON and OFF Setpoint Parameters (P334 to P337)
- Installation Records Parameter
- Range Calibration Parameters
- Rate Parameters
- Measurement Verification Parameters
- Communication Parameters
- Echo Processing Parameters
- Algorithm Parameters
- TVT Adjustment Parameters
- P837 Auto Near TVT
- Test Parameters
- _
- Communications: Modbus Register Map
- Communications: Data Types
- Error Handling
- Troubleshooting
- Maintenance(
- Appendix I
- Appendix II
- Appendix III
- Appendix IV
- Appendix V
Page 50 IQ Radar 300 PL-611
Applications
Applications
Applications
Applications
Application Example: Horizontal Tank with
Volume
Note: The minimum distance from the antenna face to the target is limited by the
near blanking, P800.
The application is to obtain a level measurement and corresponding 4-20 mA output
proportional to material levels in a chemical tank. The bottom of the antenna flange is
3.5m from the tank bottom. The empty level is 0m (bottom) and the full level (span) is
3.0m from the bottom. The maximum rate of filling or emptying is about 0.1 m/min. In the
event of a loss of echo, the IQ Radar 300 is to go into fail-safe Hi after 2 minutes.
Parameter Enter
P999
----
master reset
P001
1
mode of measurement = level
P002
1
material = liquid
P003
2
measurement response = 1m/minute
P004
240
antenna
= dielectric rod, standard
length
P005
1
units = metres
P006
5
empty distance = 5m
P007
4.5
span = 4.5m
P050
7
tank shape = parabolic ends
P051
8000
maximum volume = litres
P052
.8
tank dimension A = metres
P053
6
tank dimension L = metres
P070
2
fail-safe timer = 2 minutes
P071
1
fail-safe = Hi
P820
6
algorithm = first echo
P830
4
TVT type = factory
P837
(Note below)
2 & 1
Auto Near TVT = 0 (factory set)
Note: Only set P837 if product is at least 2.5m (78”) away. If closer, leave P837 at 0
00
0
until such time the level drops below 2.5m (78”).
Run: To start normal operation, press the PROGRAM key .
3000mm
122.4
500mm
(20.4”)