User's Manual

_
PL-550 37
PARAMETER DESCRIPTION _
P004 antenna
Identifies antenna configuration.
entry: 240 = dielectric rod
241 = rod + 50 mm extension
242 = rod + 100 mm extension
243 = rod + 150 mm extension (50 + 100mm)
P005 units
Determines the units for programming and measurement.
entry: 1 = metres
2 = centimetres
3 = millimetres
4 = feet
5 = inches
P006 empty
Distance from flange face to empty
level or maximum target range.
P007 span
Distance from empty (P006) to
full/100% level or minimum target
range.
P070 fail-safe timer
The amount of time delay, in minutes, before going into fail-safe mode.
P071 fail-safe material level
Selects the default measurement in the event that the fail-safe timer
expires.
entry: 1 = high; maximum span value
2 = low; minimum span value
3 = hold; hold current value
P200 mA range
Enables the mA output function by selecting the range and
relationship to span.
Refer to Functional / mA output.
entry: 1 = 0 to 20 mA
2 = 4 to 20 mA
3 = 20 to 0 mA
4 = 20 to 4 mA