Operating instructions
6 OXOR-E · Supplementary Operating Instructions · 8013035 V2.0 · © SICK AG
Important Information
Subject to change without notice
1.1 Service life of the oxygen sensor
The Analyzer module OXOR-E uses an electrochemical cell as oxygen sensor. The electro-
chemical cell has a limited service life and probably needs to be replaced several times
during the overall operating time (detailed information → p. 24, §5.2).
1.2 Application limitations
Usage
Unfavorable sample gas compositions, e.g. aerosols or high SO
2
concentrations can
shorten the service life of the electrochemical cell (→ p. 24, §5.2).
Assembly
Operate oxygen sensors upright (see “UP” marking).
▸
Assemble the S800 enclosure so that the base of the enclosure is horizontal.
1.3 Additional documentation/information
This document supplements the Operating Instructions for GMS800 gas analyzers. It
extends the “GMS800“ Operating Instructions with technical information on the OXOR-E.
▸
Observe the Operating Instructions delivered with the “GMS800”.
▸
Whenever possible, keep the oxygen sensor upright even when stored as a
spare part.
The “GMS800” Operating Instructions also specify all further documents
belonging to the individual device.
NOTICE:
▸
Pay primary attention to any individual information provided.