1.1 Service life of the oxygen sensor
The Analyzer module OXOR-E uses an electrochemical cell as oxygen sensor. The electro-
chemical cell has a limited service life and probably needs to be replaced several times
during the overall operating time (detailed information p. 24, §5.2).
1.2 Application limitations
Usage
Unfavorable sample gas compositions, e.g. aerosols or high SO
2
concentrations can
shorten the service life of the electrochemical cell ( p. 24, §5.2).
Assembly
Operate oxygen sensors upright (see “UP” marking).
Assemble the S800 enclosure so that the base of the enclosure is horizontal.
1.3 Additional documentation/information
This document supplements the Operating Instructions for GMS800 gas analyzers. It
extends the “GMS800“ Operating Instructions with technical information on the OXOR-E.
Observe the Operating Instructions delivered with the “GMS800”.
Whenever possible, keep the oxygen sensor upright even when stored as a
spare part.
The “GMS800” Operating Instructions also specify all further documents
belonging to the individual device.