Operating instructions

Technical Data
OXOR-E · Supplementary Operating Instructions · 8013035 V2.0 · © SICK AG 29
Subject to change without notice
6.3 Technical gas specifications
6.4 Materials with sample gas contact
6.5 Measuring ranges
6.6 Approvals
6.7 Auxiliary power supply for the module
Allowable sample gas temperature:
[1]
[1] Keep constant during operation.
0 45 °C (32 113 °F)
[2]
[2] When a sample gas cooler is used: Always above the cooler temperature (dew point).
Allowable sample gas dew point: Below ambient temperature
Particles in the sample gas: Free from dust and aerosols
[3]
[3] When entering the gas analyzer.
Allowable sample gas pressure
[4]
[4] Relative to the ambient/atmospheric air pressure.
For gas paths with hoses: –200 +300 hPa (–0.2 +0.3 bar)
For gas paths with pipes: –200 +1000 hPa (–0.2 +1.0 bar)
Sample gas volume flow
[1]
Minimum: 5 l/h (83 cm
3
/min)
Maximum: 100 l/h (1660 cm
3
/min)
[5]
[5] In potentially explosive atmospheres: Observe approval requirements.
With built-in gas pump:
[6]
[6] Option in Gas module.
30 60 l/h (500 1000 cm
3
/min)
Standard: 30 l/h (500 cm
3
/min)
Component Material
Measuring cell
Viton B, PVDF, stainless steel 1.4571, FEP
Base
Measuring component Measuring range
Technical Performance-tested
[1]
[1] Approvals §6.6
O
2
10 % by vol. 25 % by vol.
25 % by vol.
Conformities OXOR-E
EN 15267-3
EN 14181
2000/76/EC (17th BImSchV)
2001/80/EC (13th BImSchV)
27th BImSchV
Voltage supply: 24 VDC
Power input: 5W