5.1 Information on shutting down
5.2 Service life of the oxygen sensor
Limited service life
The electrochemical reaction slowly uses up the electrolyte in the oxygen sensor. The
oxygen sensor must therefore be renewed in certain intervals.
Unfavorable sample gas compositions can shorten the normal service life. Example:
Aerosols, high SO
2
concentrations.
Recommended maintenance interval
As preventative measure, renew the oxygen sensor after about two years operating time
( p. 25, §5.4).
Criteria for the end of the service life
Reaction time of the O
2
measurement slowly increases.
Reference point drift for O
2
increases quickly, i.e. O
2
sensitivity decreases rapidly.
5.3 Spare parts
The electrochemical cell is used up when in contact with air even when it is not
in operation.
When the gas analyzer is shutdown or stored (recommendation): Close off
the sample gas path of the gas analyzer gas-tight to prevent contact with
ambient air.
Drift is checked automatically during adjustments ( p. 20, §3.4.4).
Part No. Description Comprises
2054673 Spare parts set, oxygen sensor Oxygen sensor (with sealing ring)
( Fig. 4)
Locking varnish
[1]
[1] For fixing screw
2048615 Spare part set, OXOR-E Oxygen sensor
Holder (base) for oxygen sensor
PVDF screw-in joint for hose 6/4 mm
PVDF screw plug G1/8"
Fixing screw
Locking varnish
[1]
Whenever possible, store the oxygen sensor upright in a cool location (see
marking “UP” Fig. 4).
Store the oxygen sensor packed airtight or keep the opening in the connec-
tion gas-tight (as delivered).
Maintain the allowable storage temperature: –20 +60 °C.
Longer storage periods shorten the oxygen sensor service life.
Do not store the oxygen sensor as spare part for longer periods.