Operating instructions

22 OXOR-E · Supplementary Operating Instructions · 8013035 V2.0 · © SICK AG
Adjustment Information
Subject to change without notice
4.1 Setting parameters and controlling adjustments
The control unit controls the adjustments.
Individual adjustment of each shown measuring component and each measuring
range.
Programming of the adjustment parameters for each measuring component of the
GMS800 Technical Information “Basic Control Unit (BCU)”
Manual start of an adjustment procedure Operating Instructions of the control unit
4.2 Adjustment interval
Adjust the Analyzer module OXOR-E in regular intervals. Recommendation: Weekly.
General information concerning purpose, prerequisites and frequency of adjustments
“Series GMS800” Operating Instructions
4.3 Simplifications for adjustments
Zero point adjustments, omitting
Zero point adjustment is not required due to the characteristics of the electrochemical cell.
Zero point adjustments for O
2
are not necessary when measuring component O
2
is
measured with the Analyzer module OXOR-E.
Reference point adjustments with air
Air can be used for reference point adjustment for the O
2
measurement when the end
value of the O
2
measuring range is at least 21 percent by volume.
Basic information on the test gases Operating Instructions “Series GMS800”
This allows you to use air for zero point adjustment of all other measuring com-
ponents of the GMS800 (as long as no metrological or physical reasons speak
against using air). Just make sure that air is not used for a zero point adjust-
ment for O
2
.
You need just air as span gas for routine adjustments when your GMS800 is
not only fitted with the Analyzer module OXOR-E but also the UNOR/MULTOR
Analyzer module and this is fitted with the adjustment unit (option).
Use the air for zero point adjustment of the UNOR/MULTOR measuring compo-
nents and for reference point adjustment of measuring component O
2
. Use the
adjustment unit for reference point adjustment of the UNOR/MULTOR measur-
ing components.