Specifications

NOTES
. To stop the calibration operation, press any key (Ex. RE, C/AC, T key) without
pressing the MODE key to return to the working area.
. O-Err signs that the reference weight is over the full capacity.
. The calibration is available with a reference weight of over 50 of the scale
capacity. Nevertheless, we recommend to adjusting by a full capacity weight.
1-Err will appear if the load is less than 50 of the capacity.
. 2-Err will appear if the error exceeds 1 of the capacity, or any other object than
a reference weight. Or, perhaps the scale is defective.
BATTERY OPTION Nicd battery pack
Charge hours : Approx. 12 hours ( at 120VAC use) to 15 hours (at 230VAC use)
under power switch of the scale OFF.
Operation hours : Approx. 5 hours continuous operation without output.
Auto-Sleep : Sleeps 3 minutes after last operation, for power saving.
Low Battery : LOW-BATT symbol appears on when the voltage falls under
the specification.
20