Instruction manual

29
7. Troubleshooting
7. Troubleshooting
Check the following points before calling for servicing.
Symptom Cause ( remedy)
The power lamp does not light up when
the power switch is turned ON.
AC adapter is not connected.
AC adapter is not receiving power.
The batteries are out.
No data can be received from the balance. The connection cable is disconnected. ( Switch off
EP-90 and the balance, then connect the cable properly.)
The balance is not active for sending data, e.g., the
balance is displaying a menu.
DIP switch settings have not been made effective. ( Set
DIP switch while power is off, then turn the power on.)
Sequence of switching power of EP-90 and the balance
was wrong. ( See 3.4 Connecting the Balance)
The printed characters deviate.
No paper feed occurs.
Paper jam is occurring. ( Reload printer paper.)
No key entry can be made. Printout is in progress.
Data is being received from the balance.
No key switch on the balance works. There is a data I/O error. ( Switch off once, then switch
on again.)
The following data has been printed out.
• The test program for servicing is running. If the power
switch is turned on with FEED key held down, the test
program is initiated.