Instruction manual

13
4. Functions and Operation
4.4 Sample Number
4.4.1 Setting sample number
Sample number is counted up at each data output.
The sample number is a 10 digit integer that cannot include a decimal point.
When an integer of fewer than 10 digits is entered, zeros fill from the left to total 10
digits.
If entering an integer of more than 10 digits or a decimal point is attempted, an error
message “ERR” is printed out.
[setting procedure]
Verify function lamp
is off. If not, press FUNC key to turn it off.
Verify sample number lamp is off. If not, press No. key to turn it off.
(example)
(1) Press Reset key
.
(2) Key in sample number.
(3) Press No. key . Print-out NO.0012345678
4.4.2 Printing of sample number
While sample number lamp is on, upon printing mass data, sample number is
automatically printed as a line before mass data.
Sample number is counted up at each data output.
Toggle No. key
to turn sample number lamp on/off.
(example)
NO.0012345678
_________________49.97g
NO.0012345679
_________________49.97g
In order to print the currently set sample number, press No. key
when sample
number lamp is off.