Chapter 5 Test and Material List
5Choose other pump rate ‘middle’ or ‘high’and repeat the previous procedures.
(pump rate definition: three pump rate under adult mode: 100/150/200ml/min; neonate:
70/90/110 ml/min
6If the accurate over range, please press ‘START CAL’.
5.5.2 AG CALIBRATE
Agent>1.5%, CO2>1.5%, N2O>40%, O2>40%
1Press ‘START CAL’, then input password ‘MINDRAY’ entering ‘CALIBRATE’ menu.
Note: Make sure the AG in ‘Measure’ mode not ‘Standby’ mode before you do calibrate.
Figure 5-10
Figure 5-11
2Input each gas concentration value according to the label on the AG bottle label.
Note: If your monitor do not have O2 module, input ‘0.0’ in O2 item.
3Open AG cover, wait for the display value stabilization.
4If the display value does not accord with the input value, please press ‘CALIBRATE’ item
and exit.
AG concentration must fit the following requirements:
Agent>1.5%, CO2>1.5%, N2O>40%, O2>40%
64