Service manual

Chapter 1 About the Product
ISO (Isoflurane)
ENF (Enflurane)
SEV (Sevoflurane)
DES (desflurane)
Airway Respiration Rate (rpm: Respiration Per Minute): AwRR
Minimum Alveolar Concentration (MAC)
4 AG waveforms (CO
2
, N
2
O, O
2
, AA)
The PM-9000 Express provides the functions of audio/visual alarm, trend graphic storage and
output, NIBP measurement, alarm event identification, large font screen, defibrillator
synchronization, oxyCRG recall, drug calculation, etc.
1.3 Environment
1.3.1 Temperature
Work mode 0 40
MINDRAY CO
2
module +5 +35
Welch Allyn mainstream CO
2
module +10 +40
Microstream CO
2
module +5 +35
Artema AION AG module +10 +35
Transportation & Storage -20 60
1.3.2 Humidity
Work mode 15% 95 % (non-condensing)
Transportation & Storage 10% 95 % (non-condensing)
Atmospheric pressure 70.0kPa 106.0kPa
1.3.3 Electrical specification
100 – 240 V AC, 50/60 Hz,
Maximum input power: 140VA; fuse: T 3A
2.3 Ah 12V lead-acid rechargeable battery
Working time of fully-charged batteries in normal status: 120 minutes (2 batteries).
From the first low-battery alarm, the batteries can supply power to the patient monitor for 5
more minutes.
Maximum charging time: 12h
4.4Ah 11.1V lithium battery
Working time of fully-charged batteries in normal status: 300 minutes (2 batteries).
From the first low-battery alarm, the batteries can supply power to the patient monitor for 5
more minutes.
Maximum charging time: 6.5h
3