enter the factory calibration mode and the LCD display “ ”.
2. Set the black body to 35.00℃(95.00℉), measure for 3 times, the measurement result shall be
within 35±0.2℃(95.00 ± 0.4℉), otherwise, the thermometer is inaccurate.
3. Set the black body to 37.00℃(98.60℉), measure for 3 times, the measurement result shall be
within 37±0.2℃(98.60 ± 0.4℉), otherwise, the thermometer is inaccurate.
4. Set the black body to 39.00℃(102.00℉), measure for 3 times, the measurement result shall
be within 39±0.2℃(102.00 ± 0.4℉), otherwise, the thermometer is inaccurate.
After finishing calibration, remove the battery to exit the calibration mode.
13 Maintenance
1) We do not authorize any institutions or individuals to maintain and repair of the product. If
you suspect that the products have any questions, please contact the manufacturer or
distributor to handle the case.
2) The user must not attempt any repairs to the device or any of its accessories. Please
contact the retailer for repair.
3) Opening of the equipment by unauthorized agencies is not allowed and will terminate any
claim to warranty.
4) Cleaning/disinfecting the device after each use,detail method see the chapter 11
Cleaning/disinfection and Care
WARNING: No modification of this equipment is allowed!
14.Packing list
Quantity Parts Part No.
1pc Infrared Thermometer(main unit) TH10A
2PCS 2 AAA batteries AAA
1PC Forehead cap(Installed on main unit) TH10ASMGW
1PC Battery cover(Installed on main unit)
TH10AASDCW
1pc User Manual /
1pc Certificate of qualification /
15.Trouble-shooting
Troubles or error
message
Checklists or situation Countermeasures or solution