User's Manual
Page 32 / 35
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
E.g.: 25/30 means 25 periods at 50 Hz or 30 periods at 60 Hz.
Guidance and manufacture’s declaration – electromagnetic immunity
The upper arm blood pressure monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the upper arm blood pressure monitor should assure that it is used in such an
environment.
Immunity test
IEC 60601-1-2
test level
Compliance level
Electromagnetic environment -
guidance
Proximity fields from
RF wireless
communications
equipment
IEC 61000-4-3
See the following
table
Complies
TestspecificationsforENCLOSUREPORTIMMUNITYtoRFwirelesscommunicationsequipment
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power
(W)
Distance
(m)
Immunity
TestLevel
(V/m)
385 380–390 TETRA400
Pulsemodulation
b)
18Hz
1,8 0,3 27
450 430–470 GMRS460,FRS460
FM
c)
±5kHzdeviation1
kHzsine
2 0,3 28
710
704–787 LTEBand13,17
Pulsemodulation
b)
217Hz
0,2 0,3 9 745
780
810
800–960
GSM800/900,
TETRA800,iDEN
820,CDMA850,
LTEBand5
Pulsemodulation
b)
18Hz
2 0,3 28
870
930
1720
1700–1990
GSM1800;CDMA
1900;GSM1900;
DECT;LTEBand1,
3,4,25;UMTS
Pulsemodulation
b)
217Hz
2 0,3 28
1845
1970
2450 2400–2570
Bluetooth,WLAN,
802.11b/g/n,RFID
2450,LTEBand7
Pulsemodulation
b)
217Hz
2 0,3 28