Authorized representative in the European Community.
Batch code
SN
Serial number
Date of manufacture
Keep dry
Upward
Fragile, handle with care
Pile Limit 6 layers
14. Package list
Only use original accessories. Check that the contents of the delivery are complete.
Quantity Parts Part No.
1pc Main unit SP10A
1pc USB line 22-23
1pc Cuff(with Air hose) USB100032W
1pc User Manual /
15.Electromagnetic Compatibility
Statement:
The equipment with following ESSENTIAL PERFORMANCE is intended used in Home healthcare
environment and professional healthcare facility environment.
Essential Performance:
According to IEC 80601-2-30: 2018, we have conducted EMC testing base on the following essential
performance:
Change in the reading for the measurement of the cuff pressure at any point of the nominal
measurement range shall not be greater than 2mmHg (0.3kpa) during and after exposure to non-
transient phenomena and after exposure to transient phenomena.