Problem Solution
Button was pressed, but the monitor has no
response
Please charge the device. If there is no charging indication,
Insert the reset hole with a needle object less than 1.5mm in
diameter to reset the device.
The device frequently fails to measure the
blood pressure values, or the values
measured are too low or too high
1. check the position of the cuff, and re-wrap the cuff at the
correct position
2. Keep quiet and relaxed and re-measure the blood-pressure
Blood pressure measured differs from those
values measured by the doctor. Every
measurement produces a different value
although the instrument functions normally
and the values displayed are normal.
1. attempt to carry out the measurements regularly at the
same time of each day, since the blood-pressure changes
during the course of the day
2. record the daily development of the values and consult your
doctor.
The air pump works, but the air pressure does
not rise.
Check if the arm cuff connection is good and there is no air
leak
If the above situation cannot be excluded in use, or any failure other than the above occurs· please contact
customer service!
10.Storage
Storage environment: store in accordance with the storage environment conditions specified in this manual,
avoid moisture, high temperature, falling, dust, sun exposure, vibration, chemicals or corrosive gas, etc.
11 . Standard list
SHENZHEN YOLANDA TECHNOLOGY CO., LTD. declares that the Upper arm blood pressure monitor complies
with following regulations and normative documents/standards:
2017/745 Medical device regulation
COUNCIL DIRECTIVE 2014/53/EU of 16 April 2014 relating radio equipment (RED)
EN ISO 15233-1 Medical devices -- Symbols to be used with medical device labels,
labelling and information to be supplied -- Part 1 General
requirements
EN 1041
Information supplied by the manufacturer with medical devices
EN 60601-1
Medical electrical equipment Part 1: General requirements for basic
safety and essential performance
EN 60601-1-2
Medical electrical equipment -- Part 1-2: General requirements for
basic safety and essential performance - Collateral standard:
Electromagnetic compatibility - Requirements and tests
EN 60601-1-6
Medical electrical equipment – Part1-6: General requirements for
basic safety and essential performance – Collateral standard:
Usability
EN 60601-1-11
Medical electrical equipment – Part 1-11: General requirements for
basic safety and essential performance – Collateral standard:
Requirements for medical electrical equipment and medical electrical