User's Manual

22 / 42
Parameter Setting:
Parameters
Range
Threshold
It can be adjusted start from 2μV to 1000 μV with 1uV increment. The default is 10μV.
Pulse width
It can be adjusted start from 50μs to 450μs with 10μs increment.
Frequency
It can be adjusted start from 2Hz to 100Hz with 1Hz increment.
Ramp up
It can be adjusted start from 0 s to 5s with 0.1s increment.
Ramp down
It can be adjusted start from 0 s to 5s with 0.1s increment.
Duration time
It can be adjusted start from 1 s to 10s with 0.1s increment.
Interval time
It can be adjusted start from 0 s to 5s with 0.1s increment.
5.2.6.3 PAS Mode
1) Click icon to enter the home page of ETS/PAS mode.
2) Click PAS icon to switch to PAS mode.
3) Click Start icon to start the treatment.
4) Press +” or “-” on the stimulator to adjust the intensity, and the intensity value will display on the iPad.
5) User can click icon of “LOW”, “MED”, or “HIG” to display different range of EMG.
“LOW”: 0-300μV
“MED”: 0-600μV
“HIG”: 0-1000μV
1. Click ETS/PAS icon
2. Select PAS icon
3. Press Start icon to start the treatment