User's Manual
19 / 42
home page of NMES mode.
6) Click “Start” icon to start the treatment.
7) Press “+” or “-” on the stimulator to adjust the intensity.
1. Click NMES icon
2. Click the Prescription
3. Select a prescription
4. Send the selected prescription
5. Send successfully
6. Press Start icon to start the treatment
Prescription Selection:
The user can choose different prescription based on his/her needs. According to the international standard for electrical
stimulation, we have provided 42 preset prescription and 8 customizable prescription.
Preset prescriptions
No.
Parameters
Duration
/
Interval
(s)
Time
(min)
Ramp up
(s)
Ramp down
(s)
Frequency
(Hz)
Pulse width
(μs)
1
3/2
5
0
0
5
250
2
5/8
22
2
1
35
150
3
8/15
20
3
1
60
50
4
6/10
11
2
2
45
100
5
3/2
5
0
0
5
200
6
5/8
17
2
1
30
200
7
8/15
16
3
1
60
200
8
6/10
11
2
2
45
200
9
3/2
5
0
0
5
200
10
5/10
17
2
1
40
200