Table 1
MANUFACTURER’S DECLARATION-ELECTROMAGNETIC EMISSION
This equipment is intended for use in the electromagnetic environment specified below. The user should comply to
specifications ensuring the device is only used in an appropriate environment.
Emissions test
Compliance
Electromagnetic
Environment - Guidance
RF emissions,
CISPR 11
Group 1
This equipment uses RF energy only for its internal function. Its RF
emissions are very low and are not likely to cause any interference in
nearby electronic.
RF emissions,
CISPR 11
Class B
This equipment is suitable for domestic establishments and those directly
connected to the public low-voltage power supply network.
Harmonic emissions,
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions,
IEC 61000-3-3
Complied
Table 2
MANUFACTURER’S DECLARATION-ELECTROMAGNETIC EMISSION
This equipment is intended for use in the electromagnetic environment specified below. The user should comply to
specifications ensuring the device is only used in an appropriate environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2kV, ±4kV, ±8kV, ±15
kV air
±8 kV contact
±2kV, ±4kV,
±8kV, ±15 kV air
Floors should be wood, concrete or ceramic tile. Humidity
should be at least 30% if it is synthetic materials.
Electrical fast
transients/bursts
(EFT)
IEC 61000-4-4
±2kV
100kHz repetition
frequency
±2kV
100kHz
repetition
frequency
Main power quality should be that of a typical commercial
or hospital environment.