User's Manual
25 / 44
5.2.5.3 PAS Mode
1) Click icon to enter the home page of ETS/PAS mode.
2) Click PAS icon to switch to PAS mode.
3) Click Start icon to start the treatment.
4) Press “+” or “-” on the stimulator to adjust the intensity, and the intensity value will display on the iPad.
5) User can click icon of “LOW”, “MED”, or “HIG” to display different range of EMG.
“LOW”: 0-300μV
“MED”: 0-600μV
“HIG”: 0-1000μV
1. Click ETS/PAS icon
2. Select PAS icon
3. Press Start icon to start the treatment
4. Intensity value
5. EMG ranges
Parameter Setting:
Under PAS mode, the stimulator outputs NMES electrical stimulation at regular pulse width (200μs), frequency (18Hz). The
intensity of electrical stimulation is controlled by the patient’s spontaneous EMG intensity.
5.2.5.4 EMG Mode
1) Click icon to enter the home page of EMG mode.
2) Click Start icon to start EMG test, and the whole process will last 15 seconds.
3) User can click icon of “LOW”, “MED”, or “HIG” to display different range of EMG.
“LOW”: 0-300μV