User's Manual

13
CMRR
>60 dB
Center Frequency
50/60 Hz
Linearity and dynamic range
10mV (Peak-to-valley)
Frequency response
0.67 to 40 Hz
Gain error (Max.)
±10%
HR measurement range
30 to 250 /min
Accuracy
±
2 /min or
±
2%, whichever is greater.
Memory
Capacity
BP2: up to 100 readings for Blood Pressure data
+ 10 records for ECG data.
BP2A: up to 100 readings for Blood Pressure data
Bluetooth RF
Frequency range
2.402-2.480 GHz
GFSK Modulation
Adaptive Frequency Hopping (AFH)
Wireless Quality of Service
(QoS)
Transmission Distance: 1.5m
Transmission Time:
0.5s for one Blood Pressure reading
10s for one ECG record
Data integrity: 100%
Network topology
Point-to-Point
Band width
1Mbps
Durable period
Expected service life
3 years
7. Maintenance and Cleaning
7.1 Maintenance
To protect your product from damage, please observe the following:
Store the product and the components in a clean, safe location.
Do not wash the product and any components or immerse them in water.
Do not disassemble or attempt to repair the product or components.
Do not expose the product to extreme temperatures, humidity, dust or direct sunlight.
The cuff contains a sensitive air-tight bubble. Handle this carefully and avoid all types of straining
through twisting or buckling.
Clean the product with a soft, dry cloth. Do not use petrol, thinners or similar solvent. Spots on the
cuff can be removed carefully with a damp cloth and soapsuds. The cuff must not be washed!