User’s Manual Blood Pressure Monitor Model BP2, BP2A 1
1. The Basics This manual contains the instructions necessary to operate the product safely and in accordance with its function and intended use. Observance of this manual is a prerequisite for proper product performance and correct operation and ensures patient and operator safety. 1.1 Safety Warnings and Cautionary Advices Before using the product, please ensure that you have read this manual thoroughly and fully understand corresponding precautions and risks.
themselves. Do not store the product in the following locations: locations in which the product is exposed to direct sunlight, high temperatures or levels of moisture, or heavy contamination; locations near to sources of water or fire; or locations that are subject to strong electromagnetic influences. This product displays changes in the heart rhythm and blood pressure etc. which may have various different causes.
of the blood of the patient. Do not apply the cuff on an arm with another medical electrical equipment attached. The equipment may not function properly. People who have a severe circulatory deficit in the arm must consult a doctor before using the product, to avoid medical problems. Do not self-diagnose the measurement results and start treatment by yourself. Always consult your doctor for evaluation of the results and treatment.
conditions. All interpretations should be reviewed by a medical professional for clinical decision-making. DO NOT use this product in the presence of flammable anesthetics or drugs. DO NOT use this product while charging. Remain still while recording an ECG. The detectors of ECG have been developed and tested on Lead I and II recordings only. 2. Introduction 2.
Power On/ Off Power On:Press the button to power on. Power off:Press and hold the button to power off. Press to power on the product and press again to start measuring blood pressure. Press to power on the product and touch the electrodes to start measuring ECG. 3. Memory button Press to review historical data. 4. LED indicator Blue light is on: the battery is being charged. Blue light is off: the battery is full charged not charging 5.
charge will need 2 hours. When the battery charged fully the indicator will be blue. The product works in a very low power consumption and one charge usually works for months. On-screen battery symbols which indicate the battery status can be seen on the screen. Note: The product cannot be used during charging, and if choosing a third party charging adaptor, select one that complies with IEC60950 or IEC60601-1. 3.2 Measure Blood Pressure 3.2.1 Applying the arm cuff 1.
measurement takes about 30s. 3. The blood pressure readings will scrolling appear in the product when the measurement finished. 4. The product will automatically release the cuff gas after the measurement is over. 5. Press the button to turn off the power after the measurement, then remove the cuff. 6. Press the memory button to review historical data.
3.3 Measure ECG 3.3.1 Before using ECG Before using ECG function, pay attention to the following points in order to obtain precise measurements. The ECG electrode must be positioned directly against the skin. If your skin or hands are dry, moisten them using a damp cloth before taking the measurement. If the ECG electrodes are dirty, remove the dirt using a soft cloth or cotton bud dampened with disinfectant alcohol.
3.When the bar if fully filled, the product will show the measurement result. 4.Press the memory button to review historical data. Note: Do not press the product too firmly against your skin, which may result in EMG (electromyograph) interference. The device can store maximum 10 records for ECG data. The oldest record will be overwritten when the 11th record is coming in. Please uploading data in time.
The product can pair and transmit with the phone under the wireless coexistence environment (e.g. microwaves, cell phones, routers, radios, electromagnetic anti-theft systems, and metal detectors), but other wireless product may still interface with pairing and transmission between the phone and the product under uncertain environment. If the phone and the product display inconsistent, you may need to change the environment. 4.
6. Specifications Classifications Degree of protection against electrical shock Type BF Protection against electric shock Internally powered equipment Electro-magnetic Compatibility Group I, Class B Apply part Cuff and ECG electrode Environmental Item Operating Storage Temperature 5°C to 45°C -25°C to 70°C Relative humidity (non-condensing) 10% to 95% 10% to 95% Barometric 700hPa to 1060 hPa 700hPa to 1060 hPa Degree of dust & water resistance IP22 Drop test 1.
CMRR >60 dB Center Frequency 50/60 Hz Linearity and dynamic range 10mV (Peak-to-valley) Frequency response 0.67 to 40 Hz Gain error (Max.) ±10% HR measurement range 30 to 250 /min Accuracy ±2 /min or ±2%, whichever is greater. Memory BP2: up to 100 readings for Blood Pressure data Capacity + 10 records for ECG data. BP2A: up to 100 readings for Blood Pressure data Bluetooth RF 2.402-2.480 GHz Frequency range GFSK Modulation Adaptive Frequency Hopping (AFH) Transmission Distance: 1.
• Do not drop the instrument or treat it roughly in any way. Avoid strong vibrations. • Never open the product! Otherwise the manufacturer calibration becomes invalid! 7.2 Cleaning The product can be repeatedly used. Please clean before reuse as follow: • Clean the product with a soft, dry cloth with 70% alcohol. • Do not use petrol, thinners or similar solvent. • Clean the cuff carefully with cloth soaked 70% alcohol. • The cuff must not be washed.
The device has been evaluated to meet general RF exposure requirement. The device can be used in portable exposure condition without restriction. 9. Electromagnetic Compatibility The product meets the requirements of EN 60601-1-2. Warnings and Cautionary Advices Using accessories other than those specified in this manual may result in increased electromagnetic emission or decreased electromagnetic immunity of the equipment.
The Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the Blood Pressure Monitor should assure that it is used in such an environment. Compliance Electromagnetic Immunity test IEC60601 test level level environment - guidance Floors should be wood, ± 8kV contact ± 8kV contact concrete or ceramic tile.
6Vrms 150 kHz to 80 MHz outside ISM bands should be used no closer to any part of the system, including cables, than the recommended separation distance calculated from the equation appropriate for the frequency of the transmitter. Recommended separation distances: 3.5 d P V1 3.5 d P E1 80MHz to 800MHz 7 d P E1 Radiated RF IEC61000-4-3 10V/m 80MHz to 2.7 GHz 10 V/m 800MHz to 2.
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Blood Pressure Monitor is used exceeds the applicable RF compliance level above, the Blood Pressure Monitor should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Blood Pressure Monitor.
Shenzhen Viatom Technology Co., Ltd. 4E,Building 3, Tingwei Industrial Park, No.6 Liufang Road, Block 67, Xin’an Street, Baoan District, Shenzhen 518101 Guangdong China www.viatomtech.com info@viatomtech.