The device is intended for use in the specified electromagnetic environment. The customer
or the user of the device should assure that it is used in such an environment as described
below.
Immunity
test
IEC60601
test level
Complian
ce level
Electromagnetic environment - guidance
Conduced
RF
IEC61000-4-
6
3 Vrms
150 kHz to
80 MHz
outside
ISM bands
3 Vrms
150 kHz to
80 MHz
outside
ISM bands
Portable and mobile RF communications
equipment should be used no closer to any
part of the system, including cables, than
the recommended separation distance
calculated from the equation appropriate for
the frequency of the transmitter.
Recommended separation
distances:
Pd 2.1
Radiated RF
IEC61000-4-
3
3 V/m 80
MHz to
2.5 GHz
3 V/m 80
MHz to
2.5 GHz
Recommended separation distances:
80 MHz
800 MHz:
Pd 2.1
800MHz-2.5GHz:
Pd 3.2
Where, P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
a
, should be less than the compliance
level in each frequency range
b
.
Interference may occur in the vicinity of
equipment
marked with the
following
symbol:
Note 1: At 80 MHz to 800 MHz, the separation distance for the higher frequency range
applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the device is used
exceeds the applicable RF compliance level above, the device should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the device.
b
Over frequency range 150kHz to 80MHz. For Resp field strength should be less than
1V/m.