User's Manual

16
Temperature measurement
Measurement method
Direct mode
Measurement site
Mouth,Axilla,Rectum
REFERENCE BODY
SITE
Mouth,Axilla,Rectum
Temperature Unit
Celsius(℃or Fahrenheit(℉
Temperature range
32.0 43.9℃89.6-111.0℉
Measurement accuracy
Less than 35.8℃, ±0.3℃.(Less than 96.4℉,
±0.5℉.)
35.8℃ to less than 37℃,±0.2℃.(96.4℉ to
less than 98.0℉,±0.3℉.)
37.0℃ to 39.0℃, ±0.1℃.(98.0℉ to
102.0℉, ±0.2℉.)
Greater than 39.0℃ to
41.0℃,±0.2℃.(Greater than 102.0℉ to
106.0℉,±00.3℉.)
Greater than 41.0℃,±0.3℃.(Greater than
106.0℉,±0.5℉.)
Pulse Rate measurement
Pulse Rate range
40 to 200 bpm
Pulse Rate accuracy
±2 bpm or ±2%, whichever is greater
Bluetooth RF
Frequency range
2.402 2.480 GHz
Max RF power
-10 dBm