Specifications
Table Of Contents
B-6
Guidance and manufacturer’s declaration – electromagnetic immunity
The model Dynamic ECG recorder is intended for use in the
electromagnetic environment specified below. The customer or the user o
The model Dynamic ECG recorder should assure that it is used in such a
environment.
Immunity test
IEC 60601 test
level
Comp liance
level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD)
IEC61000-4-2
± 8 kV contact
± 2 kV
,
± 4 kV
,
± 8
kV
,
± 15kV air
± 8 kV contact
± 2 kV
,
± 4 kV
,
± 8 kV,± 15kV
air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/ burst
IEC61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
n.a.
n.a.
Surge
IEC61000-4-5
± 1 kV line to line
±2 kV line to earth
n.a.
n.a.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC61000-4-11
0% U
T
0,5cycle
At
0°,45°,90°,135°,18
0°,225°,270°and
315°,
0% U
T
1cycle
and 70% U
T
25/30 cycles
Single phase:at 0°
n.a.
n.a.
Powerfrequency
(50/60 Hz)
magnetic field
IEC61000-4-8
30A/m, 50/60Hz
30A/m,50/60H
z
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE : U
T
is the AC mains voltage prior to application of the test level.
The model Dynamic ECG recordershould be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating The model Dynamic ECG recorder
d Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
n