User's Manual

Results
Fingertip Oximeter
User Manual
(Model: PF10AWPF10AW1PF10BWPF10BW1)
Download the APP software
You can view the measurements and record
list in the App. Scan the below QR code to
download the APP software for iOS and
Android system.
Instructions for Safe Operation
Make sure that there is no visible damage that may affect users
safety or measurement performance with regard to sensors and
clips. It is recommended that the device should be inspected
minimally before each use. If there is obvious damage, stop using
the device.
Special attention should be paid while the oximeter is used
constantly under the ambient temperature over 37°C, burning hurt
may occur because of overheating of the sensor at this situation.
Necessary maintenance must be performed only by qualified
service technicians. Users are not permitted to service this device.
The oximeter must not be used with devices and accessories not
specified in User Manual.
Warnings and Cautions
Explosive hazardDO NOT use the Oximeter in environment with
inflammable gas such as some ignitable anesthetic agents.
DO NOT use the Oximeter while the patient is under MRI or CT
scanning. This device is NOT MRI Compatible.
Discomfort or pain may appear if using the Oximeter continuously
on the same location for a long time, especially for patient with
poor microcirculation. It is recommended that the Oximeter should
not be applied to the same location for longer than 2 hours. If any
abnormal condition is found, please change the position of
Oximeter.
The light (the infrared light is invisible) emitted from the device is
harmful to the eyes. Do not stare at the light.
The Oximeter is not a treatment device.
Local laws and regulations must b e followed when disposing of the
device.
Keep the Oximeter away from dust, vibration, corrosive substances,
explosive materials, high temperature and moisture.
The device should be kept out of the reach of children.
If the oximeter gets wet, please stop using it and do not resume
operation until it is dry and checked for correct operation. When it
is carried from a cold environment to a warm and humid
environment, please do not use it immediately. Allow at least 15
minutes for Oximeter to reach ambient temperature.
DO NOT operate the button on the front panel with sharp
materials or sharp point.
DO NOT use high temperature or highpressure steam disinfection
on the Oximeter. Refer to Chapter 8 for instructions regarding
cleaning and disinfection.
Pay attention to the effects of lint, dust, light (including sunlight),
etc.
FCC Rules
This device complies with Part 15 of the FCC Rules. Operation is
subject to the following two conditions: (1) this device may not cause
harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation.
Note: This equipment has been tested and found to comply with the
limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause
harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from
that to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
Note: The Grantee is not responsible for any changes or modifications
not expressly approved by the party responsible for compliance. such
modifications could void the user's authority to operate the
equipment.
The device has been evaluated to meet general RF exposure
requirement.
This equipment complies with FCC's RF radiation exposure limits set
forth for an uncontrolled environment. This device and its antenna(s)
must not be colocated or conjunction with any other antenna or
transmitter.
1Overview
1.1 Intended Use
This Fingertip Oximeter is intended for measuring the pulse rate and
functional oxygen saturation (SpO
2
) through a patient’s finger. It is
applicable for s potchecking SpO
2
and pulse rate of adult and
pediatric patients in homes and medical clinics.
1.2 Views
Figure 1 Front and Rear Views
1.3 Features
Display SpO
2
, PR, PI, and Plethysmogram
Auto power On/Off
Change between PR and PI
Overlimit indication and sound
Mute sound
Four direction display
Setting menu (including overlimit setting)
Pulse beep
Wireless function (PF10AW, PF10AW1, PF10BW and PF1 0BW1 only)
Continuous or spot check measuring mode
Record list
2Charging
Charge the battery before using.
Connect th e device to computer USB or USB charging adapter with
USB cable.
: Fully charged.
: The filled part represents the remaining power. If the filled part
moves from left to right, the device is charging.
: Low battery. Please charge the device
Note: Please use the accessories that are original or approved by our
company.
3POWERON/OFF
POWER ON:
Wear the device, it will turn on automatically.
POWER OFF:
Take the device off.
It will turn off automatically after 2 seconds.
On the menu interface, if there is no key operation for about 30
seconds, the device will automatically exit the menu and then
shut down.
On the recording and playback screen, if there is no key
operation for 6 seconds, the device will automatically shut
down.
4Start/StopMeasuring
1. Open the clip and put finger inside the clip (make sure the finger is
in the correct position), and then release the clip.
2. Wait for 2 seconds, the
oximeter will power on
and start to measure.
3. The display screen shows
the measurement.
4. Get the finger out, and the d evice
will automatically power off.
Attentions for measuring:
Do not shake the finger and relax during measurement.
Do not put wet finger directly into sensor.
Avoid placing the device on the same limb which is wrapped with a
cuff for blood pressure measurement or during venous infusion.
Do not let anything block the emitting light from device, i.e. do not
use finger nail polish/paints.
Existence o f high intensive light sources, such as fluorescence light,
ruby lamb, infrare d heater or strong sunshine, etc. may cause
inaccuracy of measurement result. Please put an opaque cover on
the sensor or change the measuring site if necessary.
Vigorous exercise and electrosurgical device interference may
affect the measuring accuracy.
Nail polish may affect the measuring accuracy, and too long
fingernail may cause failure of measurement or inaccurate result.
If the first reading appears with poor waveform (irregular or not
smooth), then the reading is unlikely true, the more stable value is
expected by waiting for a while, or a restart is needed when
necessary.
If the measurements over the limits, there is a reminder sound. You
can press the Display key to mute it, or wait for 10 seconds till the
sound disappears b y itself.
5Screen
5
.1 Indications and Icons
Measuring Screen with PR
Measuring Screen with PI
Icon : indicates the wireless connection is set up between
the mobile device and oximeter (
PF10AW, PF10AW1, PF10BW and
PF10BW1 only).
Status of
Definition
Flashing in
blue
The oximeter is connecting with the mobile
devices.
Blue on
The connection between the oximeter and
mobile devices is established.
No display of
”icon
The oximeter fails to set up wireless
connection with mobile device within 3
minutes.
Hardware failure of wireless function.
Icon : indicates the countingdown time if the oximeter
works at Spot check mode. The total measuring time for Spot
check mode is 30 seconds.
5.2 Four Directions of the Screen
The oximeter supports to show the screen in four directions. Ashort
pressing of the Display Key can change display d irection by 90 ° ,
and change PR/PI at the same time. The four display directions are as
shown below.
(a) (b)
(c) (d)
For display screens of figure (b) and (d), the PI% display value
will be replaced with PR display value after 20 seconds if no key
operation.
Thedisplaydirectionisrememberedateachstartup,itwill
display the screen layout (display direction) from the last time it
was used.
5.3MeasurementEndScreen(SpotCheckMode)
When the measurement ends up for Spot check mode, the measured
SpO
2
, PR value and the analysis result of pulse rhythm will be
displayed on the screen, as shown below.
6MenuSetup
During measuring, long pressing Display key can enter the setup
menu screen.
Menu
Menu operating procedures:
1. Shortly press Display Key to choose the setting item;
2. Long press Display Key to active the setting item, then shortly
press it to modify the setting parameter;
3.
Long press Display Key to confirm the modification and exit
from this setting item.
4.
Move the setting item to “Exit, and long pressing Display Key to
exit from the setup menu.
Menu settings:
Overlimit settings: If the SpO
2
or PR value is over the defined
limits, the value will flash.
“Beep” Pulse beep option. If it is set to on, every pulse beat
makes a beep.
“Mode”: Set the measuring mode. “Continuous” and “Spot
check” for optional, the default is “Spot check”.
Spot check mode: the measuring time lasts 30 seconds
with a countingdown indication. The SpO
2
and PR
readings will freeze at the end of 30 seconds, the analysis
Multi-function
connector
Display key
Display screen
Perfusion index
Perfusion
symbol
Finger Placement
Pulse symbol
SpO
2
value
PR s
y
mbol
Pulse s trength
b
ar
PR value
Pleth wave
SpO
2
symbol
Wireless icon (PC-60FW only)
Time
countdown
Battery voltage
SpO
2
alm
Lo 89
PR alm Hi 100
PR alm Lo 30
Setting menu >>
Mode Continuous
Beep On
Display Always
Setting menu >>