Min I(Smart) User Manual
8
9. Attention
1.The Mini I(Smart I) Electric Oral Anesthesia Injector is only intended for local
anesthesia in Dentistry. It provides a constant speed pressure boost. The selection of
anesthetic and needle shall be purchased according to doctors’ clinical requirements.
Any results caused by misuse of anesthetic and needle have nothing to do with this
product.
2. To keep the handpiece clean, we recommend using a disposable protective sleeve.
The unit is delivered with disposable sleeves to be replaced prior to use for each
patient. After using, clean the handpiece and charger with clean, dry tissue.
3.The booster tube can be autoclave sterilized:
It is recommended to use type B autoclave for sterilization (sterilization parameters:
sterilization temperature: 134℃ (273°F) ±2℃.Sterilization time: not less than
4min.Therecommended sterilization period :Sterilize immediately after
use .Attention:After sterilization, check whether there is deformation, damage, crack
and other damage, if there is damage, should stop using.
4. Store the lithium ion battery in a cool, dry and well-ventilated environment,
avoiding exposure to heat. To prevent the over discharge of battery, complete a
minimum of one full charge every 1 months.
5. The handpiece battery is not removable, and must be charged with the matching
battery charger.
6. Storage requirements:
a. Temperature between -20°C ~ +55°C
b. Relative humidity less than 95%
c.Air pressure: 700 hPa ~ 1060 hPa (1 ATM = 1013.25 hPa)
d. Avoid contact with corrosive materials
7. Work environment
Ambient temperature: 5 ℃ ~ 40 ℃
Relative humidity: ≤80%
Air pressure: 700 hPa ~ 1060 hPa (1 ATM = 1013.25 hPa)
6. While in use, avoid high magnetic fields, alternating current equipment, and
electromagnetic interference impacting normal use.
8. While transporting and using the product, avoid strong vibration, drop, and shock.
9. If any malfunction of the device, please contact our company for maintenance
instead of disassembling on your own, this would void the warranty.
10. When this product must be disposed of, follow the relevant regulations of your
country.
11.Do not modify this equipment without authorization of the manufacturer