Monitoring with the TD60 at the CS Physiological Alarms
TMS60 Operator’s Manual 9 - 9
In addition, the activation state of some arrhythmias can be set as a whole with the fol-
lowing buttons that are at the bottom of the [Arrhythmia Alarms] tab.
Irr.Rhythm X
X
Off Off Off Off
Pauses/min X
X
X X On Off Off Off
Parameters
Alarm levels*
Activation
State
Factory default alarm
responses
High Medium Low Message
Record
on
Alarm
Print
on
Alarm
Paging
Switch
* X indicates available alarm level, — indicates alarm level not available
** The factory default settings are in bold.
NOTE
The priority of lethal arrhythmia alarms is always high. It is unchange-
able.
Button Description
Lethals Only Sets the lethal arrhythmia alarms to on and all non-lethal arrhythmia alarms to off.
All Alarm On Sets all arrhythmia alarms to on.
All Alarm Off Sets all arrhythmia alarms to off.
This button is enabled when the [Lethal Arrh. OFF] option in the [Tel emet ry] tab
from the [Admin Setup] menu is set to [Enable].