Physiological Alarms Monitoring with the TD60 at the CS
9 - 4 TMS60 Operator’s Manual
9.2.2 Parameter Alarm Responses
At the CS, the following parameter alarm responses may occur when a parameter alarm
occurs.
Record on Alarm: directs the CS to send the alarm data to the configured recorder.
Print on Alarm: directs the CS to send the alarm data to the configured printer.
Paging Switch: directs the CS to send an alarm to the beeper to notify the clinician.
SpO
2
Desat
Unit of measure: %
Invalid data: “--”
High limit N/A N/A
Low limit 0 to 100
Default: 80
0 to 100
Default: 80
PR
Unit of measure: bpm
Invalid data: “--”
High limit (Low limit + 2) to 300
Default: 120
(Low limit + 2) to 300
Default: 160
Low limit 18 to (high limit - 2)
Default: 50
18 to (high limit - 2)
Default: 75
Parameters Alarm limit Adult Pediatric
* ΔQTc is only available for Mindray ECG algorithm.
NOTE
The SpO
2
Desat alarm limit is restricted such that the alarm limit can not
be higher than the SpO
2
low limit.