User's Manual
Table Of Contents
- 1 Safety
- 2 General Product Description
- 3 Getting Started
- 3.1 Unpacking and Checking
- 3.2 Environmental Requirements
- 3.3 Connecting the ECG Leadwire
- 3.4 Installing the Batteries
- 3.5 Powering On the Unit
- 3.6 Understanding Touch Gestures
- 3.7 Basic Operations
- 3.7.1 Understanding the Screen Display Orientation
- 3.7.2 Browsing the Screen Display
- 3.7.3 Switching the Screen Display Orientation
- 3.7.4 Flipping the Landscape Display
- 3.7.5 Displaying the Quick Keys Area
- 3.7.6 Entering the Main Menu
- 3.7.7 Turning the Display Off
- 3.7.8 Turning the Display On
- 3.7.9 Unlocking the Screen
- 3.7.10 Acknowledging the Nurse Call
- 3.8 Using the Pouch
- 4 User Configurations
- 5 Patient Management
- 6 Alarms
- 7 Monitoring ECG
- 7.1 Introduction
- 7.2 Safety
- 7.3 Preparation for Monitoring ECG
- 7.4 Changing the ECG Settings
- 7.5 Understanding the ECG Display
- 8 Monitoring SpO2 (Optional)
- 9 Monitoring with the TD60 at the CS
- 9.1 Introduction
- 9.2 Physiological Alarms
- 9.3 ECG Monitoring
- 9.4 QT Monitoring
- 9.5 ST Monitoring
- 9.6 Arrhythmia Monitoring
- 9.7 SpO2 Monitoring
- 10 Configuring the TD60
- 10.1 Introduction
- 10.2 Maintenance Menu
- 10.2.1 Entering the Maintenance menu
- 10.2.2 Configuring the General Menu
- 10.2.3 Configuring the Alarms Menu
- 10.2.4 Quick Keys Menu
- 10.2.5 Configuring the Defaults Menu
- 10.2.6 Transferring a Configuration
- 10.2.7 Screen Lock Menu
- 10.2.8 Changing the Passwords
- 10.2.9 Changing the Device Name
- 10.2.10 Demo Mode
- 10.2.11 Service Menu
- 11 Battery
- 12 Troubleshooting
- 13 Cleaning and Disinfecting
- 14 Maintenance
- 15 Accessories
- A Product Specifications
- B EMC
- D Symbols and Abbreviations
- E Anomaly
Monitoring with the TD60 at the CS Physiological Alarms
TMS60 Operator’s Manual 9 - 3
9.2.1 Factory Default Parameter Alarm Limits
The following table lists the factory default alarm limits for all parameters.
WARNING
• If you switch off all arrhythmia alarms, the CS cannot give any arrhyth-
mia alarms. Always keep the patient under close surveillance.
Parameters Alarm limit Adult Pediatric
HR
Unit of measure: bpm
Invalid data: “--”
High limit (Low limit + 2) to 300
Default: 120
(Low limit + 2) to 350
Default: 160
Low limit 15 to (high limit - 2)
Default: 50
15 to (high limit - 2)
Default: 75
ST Single
Unit of measure: mV
Invalid data: “--”
High limit (Low limit + 0.2) to 2.00
Default: 0.2
(Low limit + 0.2) to 2.00
Default: 0.2
Low limit -2.00 to (high limit – 0.2)
Default: -0.2
-2.00 to (high limit – 0.2)
Default: -0.2
ST Dual
Unit of measure: mV
Invalid data: “--”
High limit (Low limit + 0.2) to 2.00
Default: 0.2
(Low limit + 0.2) to 2.00
Default: 0.2
Low limit -2.00 to (high limit – 0.2)
Default: -0.2
-2.00 to (high limit – 0.2)
Default: -0.2
QTc
Unit of measure: ms
Invalid data: “--”
High limit 200 to 800
Default: 500
200 to 800
Default: 480
Low limit N/A N/A
∆QTc*
Unit of measure: ms
Invalid data: “--”
High limit 30 to 200
Default: 60
30 to 200
Default: 60
Low limit N/A N/A
SpO
2
Unit of measure: %
Invalid data: “--”
High limit (Low limit + 1) to 100
Default: 100
(Low limit + 1) to 100
Default: 100
Low limit 0 to (high limit - 1)
Default: 90
0 to (high limit - 1)
Default: 90
* ΔQTc is only available for Mindray ECG algorithm.